Experimental measurements of micromotions of the cementless intervertebral disc prostheses in the cadaver bone

      Highlights

      • Two groups were compared depending on the anchoring (spikes vs. fin).
      • Two groups were compared regarding the bearing concept (non/semi- vs. constrained).
      • Majority of the implants showed values clearly below 200 μm in all planes.
      • Implants with spikes anchoring und non-constrained concept might more adventurous.

      Abstract

      Background

      Sufficient primary stability is mandatory for successful bony prosthetic incorporation. Therefore, defined micromotion values of 150 μm should not be exceeded as higher values might compromise the ingrowth of bone trabeculae to the implant surface. The aim of this study was to evaluate the primary stability of different cementless disc prosthesis in a cadaver model.

      Methods

      Four different implants with different anchoring and bearing concept were tested with a target level of L4/5. 26 specimens were randomly allocated to 1 of the 4 different implants with 6 speciments in each group. Two groups were formed depending on the anchoring (spikes vs. fin) and bearing concept (non−/semi- vs. constrained). Each implant was tested regarding primary stability in a hydraulic simulator allowing simultaneous polyaxial segment movements and axial loading. The measurements were recorded on the lower plate of the prosthesis.

      Findings

      The majority of the implants showed micromotion values below 200 μm in all planes. Only one prosthesis presented borderline longitudinal amplitudes that were significant higher than the other planes. Furthermore, significant differences were observed in the sagittal plane when comparing spike and keel anchoring. Spike anchoring implants showed superior tresults to keel anchoring implants (40 μm vs. 55 μm; p = .039), while the non−/semi-constrained bearing concept was more advantageous compared to constrained group (40 μm vs. 63 μm; p = .001).

      Interpretation

      Spike anchoring and non-constrained implants might provide better primary stability.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Clinical Biomechanics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Auerbach J.D.
        • Ballester C.M.
        • Hammond F.
        • Carine E.T.
        • Balderston R.A.
        • Elliott D.M.
        The effect of implant size and device keel on vertebral compression properties in lumbar total disc replacement.
        Spine J. 2010; 10: 333-340
        • Bragdon C.R.
        • Burke D.
        • Lowenstein J.D.
        • et al.
        Differences in stiffness of the interface between a cementless porous implant and cancellous bone in vivo in dogs due to varying amounts of implant motion.
        J. Arthroplast. 1996; 11: 945-951
        • Branson P.J.
        • Steege J.W.
        • Wixson R.L.
        • Lewis J.
        • Stulberg S.D.
        Rigidity of initial fixation with uncemented tibial knee implants.
        J. Arthroplast. 1989; 4: 21-26
        • Brinjikji W.
        • Diehn F.E.
        • Jarvik J.G.
        • et al.
        MRI findings of disc degeneration are more prevalent in adults with low back pain than in asymptomatic controls: a systematic review and meta-analysis.
        AJNR Am. J. Neuroradiol. 2015; 36: 2394-2399
        • Burke D.W.
        • O’Connor D.O.
        • Zalenski E.B.
        • Jasty M.
        • Harris W.H.
        Micromotion of cemented and uncemented femoral components.
        J. Bone Joint Surg. Br. Vol. 1991; 73: 33-37
        • Cleveland D.
        The use of methylacrylic for spinal stabilization after disc operations.
        Marquette Med. Rev. 1955; 20 (S 62–64)
        • Cunningham B.W.
        • Gordon J.D.
        • Dmitriev A.E.
        • Hu N.
        • McAfee P.C.
        Biomechanical evaluation of total disc replacement arthroplasty: an in vitro human cadaveric model.
        Spine. 2003; 28: S110-S117
        • Eysel P.Z.
        • Heine J.
        Die künstliche Bandscheibe.
        Dt Ärztebl. 2000; 97 (A 3092–3096)
        • Fernstrom U.
        Arthroplasty with intercorporal endoprothesis in herniated disc and in painful disc.
        Acta Chir. Scand. Suppl. 1966; 357: 154-159
        • Galbusera F.
        • Bellini C.M.
        • Zweig T.
        • et al.
        Design concepts in lumbar total disc arthroplasty.
        Eur. Spine J. 2008; 17: 1635-1650
        • Garcia J.B.
        • Hernandez-Castro J.J.
        • Nunez R.G.
        • et al.
        Prevalence of low back pain in Latin America: a systematic literature review.
        Pain Physician. 2014; 17: 379-391
        • Hisey M.S.
        • Zigler J.E.
        • Jackson R.
        • et al.
        Prospective, randomized comparison of one-level Mobi-C cervical total disc replacement vs. anterior cervical discectomy and fusion: results at 5-year follow-up.
        Int. J. Spine Surg. 2016; 10
        • Lee C.K.
        Accelerated degeneration of the segment adjacent to a lumbar fusion.
        Spine. 1988; 13: 375-377
        • Lewallen E.A.
        • Riester S.M.
        • Bonin C.A.
        • et al.
        Biological strategies for improved osseointegration and osteoinduction of porous metal orthopedic implants.
        Tissue Eng. B Rev. 2015; 21: 218-230
        • Mehren C.
        • Wuertz-Kozak K.
        • Sauer D.
        • Hitzl W.
        • Pehlivanoglu T.
        • Heider F.
        Implant design and the anchoring mechanism influence the incidence of heterotopic ossification in cervical total disc replacement at 2-year follow-up.
        Spine. 2019; 44: 1471-1480
        • Nachemson A.L.
        Disc pressure measurements.
        Spine. 1981; 6: 93-97
        • Pilliar R.M.
        • Lee J.M.
        • Maniatopoulos C.
        Observations on the effect of movement on bone ingrowth into porous-surfaced implants.
        Clin. Orthop. Relat. Res. 1986; 208: 108-113
        • Rousseau M.A.
        • Bradford D.S.
        • Bertagnoli R.
        • Hu S.S.
        • Lotz J.C.
        Disc arthroplasty design influences intervertebral kinematics and facet forces.
        Spine J. 2006; 6: 258-266
        • Schmidt H.
        • Midderhoff S.
        • Adkins K.
        • Wilke H.J.
        The effect of different design concepts in lumbar total disc arthroplasty on the range of motion, facet joint forces and instantaneous center of rotation of a L4-5 segment.
        Eur. Spine J. 2009; 18: 1695-1705
        • Shim C.S.
        • Lee S.
        • Maeng D.H.
        • Lee S.H.
        Vertical split fracture of the vertebral body following total disc replacement using ProDisc: report of two cases.
        J. Spinal Disord. Tech. 2005; 18: 465-469
        • Volz R.G.
        • Nisbet J.K.
        • Lee R.W.
        • McMurtry M.G.
        The mechanical stability of various noncemented tibial components.
        Clin. Orthop. Relat. Res. 1988; 226: 38-42
        • Wilke H.J.
        • Neef P.
        • Caimi M.
        • Hoogland T.
        • Claes L.E.
        New in vivo measurements of pressures in the intervertebral disc in daily life.
        Spine. 1999; 24: 755-762
        • Witt A.N.
        • Rettig H.
        • Schlegel K.F.
        • Hackenbroch M.
        • Hupfauer W.
        Orthopädie in Praxis und Klinik.
        1980 (Band 1(2. Auflage))