Comparison of two different Radiostereometric analysis (RSA) systems with markerless elementary geometrical shape modeling for the measurement of stem migration

      Abstract

      Radiostereometric analysis (RSA) is the gold standard of measurement for in vivo 3D implants migration.

      Background

      The aim of this study was to evaluate the in vivo precision of 2 RSA marker-based systems compared with that of marker-free, elementary geometrical shape modeling RSA.

      Methods

      Stem migration was measured in 50 patients recruited from an on-going Randomized Controlled Trial. We performed marker-based analysis with the Um RSA and RSAcore systems and compared these results with those of the elementary geometrical shape RSA.

      Findings

      The precision for subsidence was 0.118 mm for Um RSA, 0.141 mm for RSAcore, and 0.136 mm for elementary geometrical shape RSA. The precision for retroversion was 1.3° for elementary geometrical shape RSA, approximately 2-fold greater than that for the other methods. The intraclass correlation coefficient between the marker-based systems and elementary geometrical shape RSA was approximately 0.5 for retroversion. All 3 methods yielded ICCs for subsidence and varus–valgus rotation above 0.9.

      Discussion

      We found an excellent correlation between marker-based RSA and elementary geometrical shape RSA for subsidence and varus–valgus rotation, independent of the system used. The precisions for out-of-plane migration were inferior for elementary geometrical shape RSA. Therefore, as a mechanism of failure, retroversion may be more difficult to detect early. This is to our knowledge the first study to compare different RSA systems with or without markers on the implant.

      Interpretation

      Marker-based RSA has high precision in all planes, independent of the system used. Elementary geometrical shape RSA is inferior in out-of-plane migration.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Clinical Biomechanics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Atkinson G.
        • Nevill A.M.
        Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine.
        Sports Med. 1998; 26: 217-238
        • Biedermann R.
        • Krismer M.
        • Stöckl B.
        • Mayrhofer P.
        • Ornstein E.
        • Franzén H.
        Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. Einzel-Bild-Rontgen-Analyse-femoral component analysis.
        J. Bone Joint Surg. Br. 1999; 81: 266
        • Bland J.M.
        • Altman D.G.
        Statistical methods for assessing agreement between two methods of clinical measurement.
        Lancet. 1986; : 307-310
        • Bøe B.G.
        • Röhrl S.M.
        • Heier T.
        • Snorrason F.
        • Nordsletten L.
        A prospective randomized study comparing electro-chemically deposited hydroxyapatite and plasma-sprayed hydroxyapatite on titanium stems 55 hips followed for 2 years with RSA and DXA.
        Acta Orthop. 2011; 82: 13-19
        • Carlsson L.V.
        • Albrektsson B.E.
        • Freeman M.A.
        • Herberts P.
        • Malchau H.
        • Ryd L.
        A new radiographic method for detection of tibial component migration in total knee arthroplasty.
        J. Arthroplasty. 1993; 8: 117
        • Ciccotti M.G.
        • Rothman R.H.
        • Hozack W.J.
        • Moriarty L.
        Clinical and roentgenographic evaluation of hydroxyapatite-augmented and nonaugmented porous total hip arthroplasty.
        J. Arthroplasty. 1994; 9: 631-639
        • Costa-Santos C.
        • Bernardes J.
        • Ayres-de-Campos D.
        • Costa A.
        • Amorim-Costa C.
        The limits of agreement and the intraclass correlation coefficient may be inconsistent in the interpretation of agreement.
        J. Clin. Epidemiol. 2011; 64: 264-269
        • Hallan G.
        • Lie S.A.
        • Furnes O.
        • Engesaeter L.B.
        • Vollset S.E.
        • Havelin L.I.
        Medium- and long-term performance of 11,516 uncemented primary femoral stems from the Norwegian arthroplasty register.
        J. Bone Joint Surg. (Br.). 2007; 89: 1574-1580
        • Hurschler C.
        • Seehaus F.
        • Emmerich J.
        • Kaptein B.L.
        • Windhagen H.
        Accuracy of model-based RSA contour reduction in a typical clinical application.
        Clin. Orthop. Relat. Res. 2008; 466: 1978-1986
        • Hurschler C.
        • Seehaus F.
        • Emmerich J.
        • Kaptein B.L.
        • Windhagen H.
        Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting.
        J. Arthroplasty. 2009; 24: 594-606
        • Ilchmann T.
        • Franzén H.
        • Mjöberg B.
        • Wingstrand H.
        Measurement accuracy in acetabular cup migration. A comparison of four radiologic methods versus roentgen stereophotogrammetric analysis.
        J. Arthroplasty. 1992; 7: 121
        • Kaptein B.L.
        • Valstar E.R.
        • Stoel B.C.
        • Rozing P.M.
        • Reiber J.H.
        A new modelbased RSA method validated using CAD models and models from reversed engineering.
        J. Biomech. 2003; 36: 873
        • Kaptein B.L.
        • Valstar E.R.
        • Spoor C.W.
        • Stoel B.C.
        • Rozing P.M.
        Model-based RSA of a femoral hip stem using surface and geometrical shape models.
        Clin. Orthop. Relat. Res. 2006; 448: 92-97
        • Kaptein B.L.
        • Valstar E.R.
        • Stoel B.C.
        • Reiber H.C.
        • Nelissen R.G.
        Clinical validation of model-based RSA for a total knee prosthesis.
        Clin. Orthop. Relat. Res. 2007; 464: 205-209
        • Kärrholm J.
        • Herberts P.
        • Hultmark P.
        • Malchau H.
        • Nivbrant B.
        • Thanner J.
        Radiostereometry of hip prostheses: review of methodology and clinical results.
        Clin. Orthop. Relat. Res. 1997; 94–110
        • Karrholm J.
        • Gill R.H.
        • Valstar E.R.
        The history and future of radiostereometric analysis.
        Clin. Orthop. Relat. Res. 2006; 448: 10
        • Kottner J.
        • Streiner D.L.
        The difference between reliability and agreement.
        J. Clin. Epidemiol. 2011; 64 (author reply 702): 701-702
        • Lazarinis S.
        • Karrholm J.
        • Hailer N.P.
        Effects of hydroxyapatite coating on survival of an uncemented femoral stem. A Swedish hip arthroplasty register study on 4,772 hips.
        Acta Orthop. 2011; 82: 399-404
        • Patriarca A.
        • Piccioni V.
        • Gigante V.
        • Benedetto C.
        The use of intravenous immunoglobulin in sine causa or alloimmune recurrent spontaneous abortion (RSA).
        Panminerva Med. 2000; 42: 193-195
        • Prins A.H.
        • Kaptein B.L.
        • Stoel B.C.
        • Nelissen R.G.
        • Reiber J.H.
        • Valstar E.R.
        Handling modular hip implants in model-based RSA: combined stem-head models.
        J. Biomech. 2008; 41: 2912-2917
        • Rankin G.
        • Stokes M.
        Reliability of assessment tools in rehabilitation: an illustration of appropriate statistical analyses.
        Clin. Rehabil. 1998; 12: 187-199
        • Ryd L.
        Micromotion in knee arthroplasty. A roentgen stereophotogrammetric analysis of tibial component fixation.
        Acta Orthop. Scand. 1986; : 1-80
        • Selvik G.
        Roentgen stereophotogrammetric analysis.
        Acta Radiol. 1990; 31: 113-126
        • Shrout P.E.
        • Fleiss J.L.
        Intraclass correlations: uses in assessing rater reliability.
        Psychol. Bull. 1979; 86: 420-428
        • Stilling M.
        • Larsen K.
        • Andersen N.T.
        • Søballe K.
        • Kold S.
        • Rahbek O.
        The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. A study of validity and reliability.
        Acta Orthop. 2010; 81: 570-578
        • Valstar E.R.
        • Gill H.S.
        Radiostereometric analysis in orthopaedic surgery: editorial comment.
        Clin. Orthop. Relat. Res. 2006; 448: 2
        • Valstar E.R.
        • Nelissen R.G.
        • Reiber J.H.
        • et al.
        The use of roentgen stereophotogrammetry to study micromotion of orthopaedic implants.
        ISPRS J. Photogramm. Remote Sens. 2002; 56: 376
        • Valstar E.R.
        • Gill R.
        • Ryd L.
        • Flivik G.
        • Börlin N.
        • Kärrholm J.
        Guidelines for standardization of radiostereometry (RSA) of implants.
        Acta Orthop. 2005; 76: 563-572
        • Wykman A.
        • Lundberg A.
        Subsidence of porous coated noncemented femoral components in total hip arthroplasty. A roentgen stereophotogrammetric analysis.
        J. Arthroplasty. 1992; 7: 197-200