A biomechanical study of the cortex-anchorage vertebral screw

      Abstract

      Objective. To obtain a comprehensive understanding on the effect of the improvement of fixation strength and on the optimal design in various geometrical parameters of a new screw system through biomechanical analyses.
      Design. A new screw with the cortex-anchorage was designed and manufactured to improve the fixation of the instrumentation for osteoporotic spine. There were four expandable wings distributed around the screw after insertion.
      Background. Screw loosening or loss of correction caused by insufficient mechanical stability on the bone-screw interface is frequently found in osteoporotic subjects. Similarly, the removal and replacement of a screw in a revision procedure substantially decreases its mechanical fixation. Since cortex is the most rigid part in the vertebral body, emphasis on the cortex-anchorage may offer an optimal fixation of screws.
      Methods. The biomechanical evaluation that consists of the pullout test and the finite element analysis was applied to identify the stabilizing effect and the optimal design for the new screw system. In the pullout experiment, the porcine vertebral body with a hollow block of cancellous bone was proposed to simulate an osteoporotic spine. This osteoporotic model was specially simulated the degeneration and destruction of the cancellous bone in vertebrae. In the finite element analysis, the reduction of elastic modulus was used in various levels of vertebral degeneration.
      Results. Pulling screws out of vertebral bodies with a hollow block of cancellous bone, the mean pullout force was 729 (SD 159) N for the conventional screws, and 1072 (SD 179) N for the new screw system. The finite element analysis showed that the longer screw with bi-cortex fixation was the better option in reducing the bony stress and increased the stability. As the height of wings changed, the stress distributed on vertebral body indicated the lowest in fixation by a screw with the largest wings. Nevertheless, there existed a least displacement of vertebral body and moderately low stress on wings’ lateral end when assembled with the middle size wings.
      Conclusion. The stabilization function of expansive wings of the new screw system was enhanced in the osteoporotic vertebra and better than that of a conventional screw. The finite element analysis showed a middle size wing could help the screw to reduce the risk of failure and to improve the vertebral stability.Relevance
      Screw loosening or loss of correction caused by insufficient mechanical stability on the bone-screw interface is frequently found in osteoporotic subjects. From the biomechanical point of view, this study had shown that a new design of screw could improve the fixation of the instrumentation for osteoporotic spine. With further investigations that includes the clinical proof and the development of a cortex-anchorage vertebral screw may provide a valuable alternative to the spinal instrumentation for the patients with osteoporosis.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Clinical Biomechanics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Aitken J.M.
        • Armstrong E.
        • Armstrong J.B.
        Osteoporosis after oophorectomy in the mature female rat and the effect of estrogen and/or progestogen replacement therapy in its prevention.
        J. Endocrinol. 1972; 55: 79-87
        • Cook S.D.
        • Salkeld S.L.
        • Whitecloud III, T.S.
        • et al.
        Biomechanical evaluation and preliminary clinical experience with an expansive pedicle screw design.
        J. Spinal Disord. 2000; 13: 230-236
        • Decoster T.A.
        • Heetderks D.B.
        • Downey D.J.
        • Ferries J.S.
        • Jones W.
        Optimizing bone screw pullout force.
        J. Orthop. Trauma. 1990; 4: 169-174
        • Dieter G.E.
        Yielding criteria for ductile metals.
        Mechanical Metallurgy. second ed. McGraw-Hill Inc, 1976 (pp. 79–82)
        • Frost H.M.
        • Jee W.S.S.
        On the rat model of human osteopenias and osteoporosis.
        Bone Miner. 1992; 18: 227-236
        • Gilbert S.G.
        • Johns P.C.
        • Chow D.C.
        • Black R.C.
        Relation of vertebral bone screw axial pullout strength to quantitative computed tomographic trabecular bone mineral content.
        J. Spinal Disord. 1993; 6: 513-521
        • Goel V.K.
        • Kim Y.E.
        • Lim T.-H.
        • Weinstein J.N.
        An analytical investigation of spinal instrumentation.
        Spine. 1988; 13: 1003-1011
        • Kalu D.N.
        The ovariectomized rat model of postmenopausal bone loss.
        Bone Miner. 1991; 15: 175-192
      1. Kleeman, B.C., Gerhart, T.N., Hayes, W.C., 1987. Augmenting screw fixation in osteoporotic trabecular bone. In: Proceedings of the Society of Biomaterials Annual Meeting, New York

        • Koranyi E.
        • Bowman C.E.
        • Knecht C.D.
        • Jansen M.
        Holding power of orthopedic screws in bone.
        Clin. Orthop. Relat. Res. 1970; 72: 283-286
        • Korovessis P.G.
        • Deligianni D.
        • Stamatakis M.
        • Missirlis Y.
        Augmentation of anterior transvertebral screws using threaded teflon anchoring.
        J. Spinal Disord. 1998; 11: 300-306
      2. Kumono, K., Niyashita, H., 1989. Internal fixation of the lumbar spine with Cotrel–Dubousset instrumentation. In: Proceedings of the International Society for the Study of the Lumbar Spine. p. 72

        • McLain R.F.
        • Fry M.F.
        • Mosely T.A.
        • Sharkey N.A.
        Lumbar pedicle screw salvage: pullout testing of three different pedicle screw designs.
        J. Spinal Disord. 1995; 8: 62-68
        • Moran J.M.
        • Berg W.S.
        • Berry J.L.
        • Geiger J.M.
        • Steffee A.D.
        Transpedicular screw fixation.
        J. Orthop. Relat. Res. 1989; 7: 107-114
        • Nunamaker D.M.
        • Perren S.M.
        Force measurements in screw fixation.
        J. Biomech. 1976; 203: 54-57
        • Pfeifer B.A.
        • Krag M.H.
        • Johnson C.
        Repair of failed transpedicle screwing for osteoporotic spine: in vitro study of the mechanical stability.
        Spine. 1993; 18: 2240-2245
        • Polly D.W.
        • Orchowski J.R.
        • Ellenbogen R.G.
        Revision pedicle screws: Bigger, longer shims––what is best?.
        Spine. 1998; 13: 1075-1076
      3. Samuelson, W.O., Simmons, E.H., 1989. Factors affecting success in adult spinal fusion. In: Proceedings of the International Society for the Study of the Lumbar Spine. p. 83

        • Saville P.D.
        Changes in skeletal mass and fragility with castration in the rat: a model of osteoporosis.
        J. Am. Geriatr. Soc. 1969; 17: 155-169
        • Schatzker J.
        • Sanderson R.
        • Murnaghan J.P.
        The holding power of orthopedic screws in vivo.
        Clin. Orthop. 1975; 108: 115-126
        • Skinner R.
        • Maybee J.
        • Transfeldt E.
        • Venter R.
        • Chalmers W.
        Experimental pullout testing and comparison of variables in transpedicular screw fixation: a biomechanical study.
        Spine. 1990; 15: 195-201
        • Smith S.S.
        • Abitbol J.J.
        • Carlson G.D.
        • et al.
        The effects of depth of penetration, screw orientation, and bone density on sacral screw fixation.
        Spine. 1993; 18: 1006-1010
        • Steffee A.D.
        • Biscup D.O.
        • Sitkowski D.J.
        Segmental spine plates with pedicle screw fixation.
        Clin. Orthop. Relat. Res. 1986; 203: 45-53
        • Vangsness C.T.
        • Carter D.R.
        • Frankel V.H.
        In vitro evaluation of the loosening characteristics of self-tapped and nonself-tapped cortical bone screws.
        Clin. Orthop. Relat. Res. 1981; 157: 279-286
        • Wittenberg R.H.
        • Lee K.S.
        • Shea M.
        • White III, A.A.
        • Hayes W.C.
        Effect of Screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength.
        Clin. Orthop. 1993; 296: 278-287
        • Zindrick M.R.
        • Wiltse L.L.
        • Widell E.H.
        • et al.
        A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine.
        Clin. Orthop. Relat. Res. 1986; 203: 99-112