Advertisement
Research Article| Volume 17, ISSUE 5, P376-382, June 2002

Computer assisted screw insertion into real 3D rapid prototyping pelvis models

      Abstract

      Objective. Show the use of computer navigation in exact screw positioning in the different pelvic bones.
      Background. Computer assisted pedicle screw insertion in the spine is an established procedure. Screw fixation is also used in highly difficult pelvic and hip surgery (arthroplasty revision surgery and tumor surgery).
      Design. Insert as long screws as possible with computer navigation into the different bones of the pelvis and compare these results with a non-computer controlled method.
      Methods. The computer navigation was done with the system of Medivision (Oberdorf, Switzerland), the software was SurgiGATE 2.1. Optically controlled spine instruments and a special calibrated drill were used. The screw insertion with and without computer navigation took place in seven real rapid prototyping pelvis models matched by pairs. Three screws were inserted into the Os ileum, one into the Os pubis and one into the Os ischium. The length of the inserted screws were measured and compared using routine statistic methods.
      Results. The mean screw length with vs. without computer navigation was 8.9 vs. 5.7 cm in the Os ileum (P=0.0001), 6.0 vs. 4.2 cm in the Os pubis (P=0.01) and 4.3 vs. 3.9 cm in the Os ischium (not significant).
      Conclusions. The use of computer navigation allows for the insertion of longer screws into the bones of the pelvis (more exact positioning), which requires a more precise original point of entry and direction of the drill (vector).
      Relevance The insertion of fixation screws in highly difficult pelvic and hip surgery (revision arthroplasty, tumor surgery) are another field for the use of computer navigation.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Clinical Biomechanics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

      1. Nolte L.P. Ganz R. Computer-Assisted Orthopaedic Surgery (CAOS). Hogrefe & Huber, Seattle1998: 1-69
        • Langlotz F.
        • Bächler R.
        • Berlemann U.
        • et al.
        Computer assistance for pelvic osteotomies.
        Clin Orthop. 1998; 354: 92-102
        • Nolte L.P.
        • Zamorano L.J.
        • Jiang Z.
        • et al.
        Image-guided insertion of transpedicular screws.
        Spine. 1995; 20: 497-500
        • Nolte L.P.
        • Zamorano L.
        • Visarius H.
        • et al.
        Clinical evaluation of a system for precision enhancement in spine surgery.
        Clin Biomech. 1995; 10: 293-303
        • Nolte L.P.
        • Visarius H.
        • Arm E.
        • et al.
        Computer-aided fixation of spinal implants.
        J Image Guid Surg. 1995; 1: 88-93
        • Foley K.T.
        • Smith M.M.
        Image-guided spine surgery.
        Neurosurg Clin North Am. 1996; 7: 171-186
        • Laine T.
        • Lund T.
        • Ylikoski M.
        • et al.
        Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patients.
        Eur Spine J. 2000; 9: 235-240
        • Merloz P.
        • Tonetti J.
        • Eid A.
        • et al.
        Computer assisted spine surgery.
        Clin Orthop. 1997; 337: 86-96
        • Schwarzenbach O.
        • Berlemann U.
        • Jost B.
        • et al.
        Accuracy of computer assisted pedicle screw placement––An in vivo computed tomography analysis.
        Spine. 1997; 22: 452-458
        • Langlotz F.
        • Stucki M.
        • Bächler R.
        • et al.
        The first twelve cases of computer assisted periacetabular osteotomy.
        Comput Aided Surg. 1997; 2: 317-326
        • Mittelmeier W.
        • Peters P.
        • Ascherl R.
        • Gradinger R.
        Rapid prototyping. Modellherstellung zur präoperativen Planung von rekonstruktiven Beckeneingriffen.
        Orthopäde. 1997; 26: 173-279
        • Lavallée S.
        Registration for computer-integrated surgery: Methodology, state of the art.
        in: Taylor R.H. Lavallée S. Burdea G.C. Mösges R. Computer-Integrated Surgery. The MIT Press, Cambridge1996: 77-97
        • Gong J.
        • Bächler R.
        • Sati M.
        • Nolte L.P.
        Restricted surface matching: A new approach to registration in computer assisted surgery.
        in: Troccaz J. Grimson E. Mösges R. Lecture Notes in Computer Science. CVRMed-MRCAS'97. Springer, Berlin1997: 597-605