Axial preload in external fixator half-pins: a preliminary mechanical study of an experimental bone anchorage system

  • I.A. Karnezis

      Affiliations

    • Corresponding Author InformationCorresponding author.
    • University Department of Orthopaedic Surgery, Bristol Royal Infirmary, Bristol BS2 8HW, UK
  • ,
  • A.W. Miles

      Affiliations

    • Department of Mechanical Engineering, University of Bath, Bath UK
  • ,
  • J.L. Cunningham

      Affiliations

    • University Department of Orthopaedic Surgery, Bristol Royal Infirmary, Bristol BS2 8HW, UK
  • ,
  • I.D. Learmonth

      Affiliations

    • University Department of Orthopaedic Surgery, Bristol Royal Infirmary, Bristol BS2 8HW, UK

Received 29 July 1997; accepted 13 May 1998.

Abstract 

Objective. To demonstrate the effect of axial preload achieved with an experimental external fixator pin prototype on the initial stability of the pin-bone interface.

Design. An in vitro mechanical study comparing pin stability of an axially-preloaded experimental pin and a radially-preloaded conventional external fixation pin.

Background. The most common problem in clinical practice of external fixation is pin site failure. An experimental external fixator half-pin has been designed introducing axial preload as an alternative method of achieving a stable pin-bone interface.

Methods. Standardized measurements of pin insertion and removal torque comparing the experimental pin prototype and a conventional external fixator pin in a synthetic composite material and ovine tibial cortical bone.

Results. There is a positive correlation between axial preload and torque resistance of the coaxial experimental half-pin as well as significantly increased torque resistance of the axially-preloaded experimental pin over the conventional pin tested. In addition, significantly lower insertion torque of the experimental pin was observed with the applied insertion technique.

Conclusions. ‘Axial preload’ achieved with appropriately-designed half pins might represent an effective biology-sparing method of increasing the mechanical stability of the pin-bone interface in unilateral external fixator frames.

Keywords:  External fixation, Pin loosening, Bone screws, Preload

No full text is available. To read the body of this article, please view the PDF online.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0268-0033(98)00047-3

Clinical Biomechanics
Volume 14, Issue 1 , Pages 69-73, January 1999