Sagittal curvature of total knee replacements predicts in vivo kinematics

  • Oliver Kessler

      Affiliations

    • Scientific Affairs, Stryker Europe, Thalwil, Switzerland
  • ,
  • Lutz Dürselen

      Affiliations

    • Institute of Orthopaedic Research and Biomechanics, University of Ulm, 89081 Ulm, Germany
    • Corresponding Author InformationCorresponding author.
  • ,
  • Scott Banks

      Affiliations

    • Department of Mechanical and Aerospace Engineering, University of Florida, Gainesville, FL, USA
  • ,
  • Henrich Mannel

      Affiliations

    • Institute of Orthopaedic Research and Biomechanics, University of Ulm, 89081 Ulm, Germany
  • ,
  • Frédéric Marin

      Affiliations

    • Institute of Orthopaedic Research and Biomechanics, University of Ulm, 89081 Ulm, Germany

Received 7 October 2005; accepted 26 July 2006. published online 10 October 2006.

Abstract 

Background

It is known that in vivo kinematics after total knee replacement is influenced by the design of the implant. The goal of this study was to show that the sagittal curvature of two different knee prostheses differing in geometric design predicts their in vivo motion behavior.

Methods

Three-dimensional tibio-femoral displacements of two prosthesis designs (single radius vs. dual radius) were measured during knee extension under weight bearing conditions by in vivo video fluoroscopy. Finite helical axes were computed to represent the tibio-femoral motions. Angular deviation α and the spatial localization deviation δ were used to characterize the motions. Angular deviation is the angle between each incremental finite helical axis and the medio-lateral axis of the femoral component of the prosthesis. The spatial localization deviation is the distance between each finite helical axis and the center of the femoral component of the prosthesis. Statistical comparisons were performed using the median and the interquartile range of the angular deviation and the spatial localization deviation.

Findings

The single-radius design showed finite helical axes concentrated at a single axis near to the medio-lateral axis of the femoral component. The angular and spatial localization deviation of the dual radius design were larger compared to the single radius design, exhibiting finite helical axes varying between two axes.

Interpretation

Video fluoroscopy in combination with finite helical axis analysis proved to be suitable methods to evaluate the in vivo kinematical behavior of total knee arthroplasty, which can be useful for implant designers. Knowledge of in vivo kinematics can also provide surgeons with more background information about the total knee arthroplasty models they implant.

Keywords: Knee kinematics, Video fluoroscopy, Helical axis analysis, Total knee arthroplasty, Single radius

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0268-0033(06)00155-0

doi:10.1016/j.clinbiomech.2006.07.011

Clinical Biomechanics
Volume 22, Issue 1 , Pages 52-58, January 2007